Agilent E5070BU/E5071BU Option 018
Hard Disk Drive Kit
User’s Guide
Second Edition
Agilent Part No. E5070-90119
June 2005
Printed in Japan
1
General Information
This chapter describes the Kit, contents of the kit, and required tools to install the Option
018 into the Agilent E5070B/E5071B.
3
General Information
Option 018 Hard Disk Drive Kit
Option 018 Hard Disk Drive Kit
The purpose of this kit is to install the option 018 Hard Disk Drive into the Agilent
E5070B/E5071B.
Contents of the Agilent E5070BU/E5071BU Option 018 Hard
Disk Drive Kit
The kit includes the parts listed in Table 1-1. If anything is missing, contact the nearest
Agilent Technologies office.
Table 1-1
Contents of the Agilent E5070BU/E5071BU Option 018 Hard Disk Drive Kit
Description
Qty.
Upgrade Manual (This manual)
HDD Assembly
1
1
4
Chapter1
2
Option Installation
This chapter describes the procedure to install the option 018 into the Agilent
E5070B/E5071B.
5
Option Installation
Introduction
Introduction
This chapter describes the procedure to install Option 018 into the E5070B/E5071B. The
procedure consists of the following items.
•
•
•
•
•
•
Writing System Correction Data of Backup Memory to Removable HDD
Caution for replacing the spare HDD with the original HDD
Update Procedure
Installation Requirement
•
TORX Screwdriver T10
Removing Removable Hard Disk
Follow the steps below to remove a removable hard disk.
CAUTION
Before removing a hard disk, you must turn off the unit’s power and then disconnect the
power cord from the outlet.
Step 1. Unscrew the four screws that hold the cover panel (1 in Figure 2-1).
Step 2. Remove the panel.
*1
Step 4. Carefully remove the flat cable (2 in Figure 2-2 ) from the removable hard disk. Release
the cable’s connector by pressing the locking hooks at the sides of the connector.
Mounting Removable Hard Disk
Follow the steps below to mount a removable hard disk.
Step 3. Secure the cover panel in place with the four screws (1 in Figure 2-1).
*1.Flat cable with connector outfitted
6
Chapter2
Option Installation
Update Procedure
Writing System Correction Data of Backup Memory to Removable
HDD
The following is a description of the procedures to write the system correction data stored
in the backup memory to the mounted removable HDD.
Step 1. Turn on the E5070B/E5071B.
Step 2. Press
.
Step 3. Press Load Project... on the softkey menu.
Step 4. A dialog box appears to allow you to select a program to be loaded. Select
RestoreSysCorFile.vba from the D:\Agilent\Service folder, and then press Open button.
Step 5. Press
.
Step 6. RestoreSysCorFile dialog box appears (Figure 2-3). Click OK.
Figure 2-3
RestoreSysCorFile Dialog Box
CAUTION
Once the program has ended, restart the E5070B/E5071B to enable the system correction
data.
8
Chapter2
Option Installation
Update Procedure
Initial Registration of E5070B/E5071B
When you start up the E5070B/E5071B for the first time or after executing system
recovery, you need to perform the initial registration of the Windows 2000 operating
system of the E5070B/E5071B.
NOTE
NOTE
You cannot use the front panel keys during the initial registration of the E5070B/E5071B
therefore connect the mouse and the keyboard before turning on the power.
If you perform the following procedure incorrectly, a message asking you whether to return
to the previous registration screen and perform the registration appears. In this case, follow
the instruction to return to the previous registration screen.
Step 1. Turns on the E5070B/E5071B.
Step 2. The Windows 2000 Professional Setup wizard appears. Click the Next > button (Figure
2-4).
Figure 2-4
Windows 2000 Professional Setup wizard
Chapter 2
9
Option Installation
Update Procedure
Step 3. In the Windows 2000 Professional Setup dialog box, select the I accept this agreement
box and click the Next >button (Figure 2-5). Then, the Windows 2000 operating system is
restarted automatically.
Figure 2-5
Windows 2000 Professional Setup dialog box
Step 4. The Network Identification wizard appears. Click the Next > button (Figure 2-6).
Figure 2-6
Network Identification wizard
10
Chapter2
Option Installation
Update Procedure
assumes the following user has logged on to this computer box and check that agena
is in the User Name box. If not, type in agena. Finally, click the Next> button (Figure 2-7).
Figure 2-7
Network Identification Wizard dialog box (1/2)
Step 6. In the Network Identification Wizard dialog box (2/2), click the Finish button to finish the
initial registration of the E5070B/E5071B (Figure 2-8). Then, the measurement display of
theE5070B/E5071B appears.
Figure 2-8
Network Identification Wizard dialog box (2/2)
Chapter 2
11
Option Installation
Update Procedure
Calibration of the Touch Screen
When you have replaced the HDD on an E5070B/E5071B equipped with an Option 016
touch screen, you have to calibrate the touch screen. Follow the procedure described below
to calibrate the touch screen.
Step 1. Press
.
Step 2. Press Service Menu.
Step 3. Press Test Menu.
Step 4. Press Adjust Touch Screen.
The touch screen calibration screen (Figure 2-9) appears.
Figure 2-9
Touch Panel Calibration Screen
Step 5. Touch the x mark on the upper left with your finger. The mark x appears also on the lower
left, upper right, and lower right. Touch the x marks in that order with your finger.
Touching the four locations described above with your finger automatically concludes the
touch screen calibration.
NOTE
With no operation on the touch screen calibration screen for a preset time, it automatically
closes and the previous measurement screen reappears.
Caution for replacing the spare HDD with the original HDD
The RestoreSysCorFile.vba is required to write the system correction data stored in the
backup memory to the orignal HDD.
Execute the following steps to copy the RestoreSysCorFile.vba file.
Step 1. Copy the RestoreSysCorFile.vba file from the D:\Agilent\Service folder to a FD before
replacing the spare HDD with the orignal HDD.
Step 2. Set the orignal HDD.
Step 3. Confirm that the RestoreSysCorFile.vba file exists on the D:\Agilent\Service of the
original HDD.
Step 4. If the file does not exist, copy the RestoreSysCorFile.vba file from the FD to the
D:\Agilent\Service folder of the original HDD.
12
Chapter2
|